| Specifications | SUN HF2 | SUN HF3 |
|
Maximum Power Waveform: |
3.3 KW High frequency |
5 KW High frequency |
|
kV Selection Fluoroscopy: Radiography |
40 - 110 kV 45 - 110 kV |
35 - 120 kV 40 - 120 kV, anatomic memories panel |
|
mA/mAs Selection
Fluoroscopy(automatic/manual/ Radiography |
0.1 - 4 mA 50 - 30 mA |
0.1 - 6 mA, "snap-shot" 12 mA 80 - 40 mA, anatomic memories panel |
|
X-ray unit: Focus size: Anodic thermal capacity: X-ray unit thermal capacity: Collimator (optional if APX HF2): Laser centering device: |
monobloc, fixed anode 0.6/1.8 mm 40 kJ 565 kJ Leaded iris and parallel blades no |
monobloc, rotating anode 0.3/0.6 mm 222 kJ 620 kJ Leaded iris and parallel blades option |
|
TV camera:
|
CRM 86:625/525 lines,Vidicon tube or X-EYE:625 lines, CCD sensor |
CRM 86:625/525 lines,Vidicon tube or X-EYE:625 lines, CCD sensor |
| Image intensifier: |
16 cm one field, 17 cm double field, 23 cm triple field |
16 cm one field, 17 cm double field, 23 cm triple field |
| Anti-scattering grid: | R8:1 - 40 I/cm | R8:1 - 40 I/cm |
|
Video processing system and storage capacity:
DSA: |
LC 10..: High definition digital memory (944x544x12bits) real time noise reduction and edge enhancement (LIH: 1-2-5-10-200 images)
|
LC 10..: High definition digital memory (944x544x12bits) real time noise reduction and edge enhancement (LIH: 1-2-5-10-200 images) or
High definition digital memory, noise |
| TV monitor(s): |
44 cm, (17") bandwidth 40 MHz at -6 dB, low radiation, brightness remote control Option: image rotation, and "flicker-free" (100 Hz) |
44 cm, (17") bandwidth 40 MHz at -6 dB, low radiation, brightness remote control Option: image rotation, and "flicker-free" (100 Hz) |
|
Cassette holder (optional grid, cassette): |
24 x 30 cm | 24 x 30 cm |
| Power supply: |
230 VAC +/- 10%-50 ou/ or 60Hz 115 VAC +/- 10%-50 ou/ or 60Hz |
240 VAC +/- 10%-50 ou/ or 60Hz 120 VAC fluoroscopy only (radiography available with extermal line adapter) |